Holt Model HCS-1 Primary Shunt Set
...for the measurement of alternating current amplitude
|
|
This Primary Alternating Current Shunt Set consists of seven
separate shunt modules, designed to plug directly into either the
Holt Model 6B, any Thermal Transfer Voltmeter or true RMS Digital
Voltmeter. Each Shunt is a four-terminal resistor providing
optimum frequency response well beyond the audio region.
Controlled thermal characteristics ensure stable current
measurements up to 26 amperes and advanced techniques are
employed to minimize inductance and skin effects. Each module is
attractively finished in dull black and individually serial numbered
for ease of traceability.
An optional Shunt Transfer Unit is available with the HCS-1
Current Shunt Set allowing convenient interconnection of the
Shunt with the associated AC-DC current sources. This
dependable, low resistance, coaxial switch unit permits rapid
transfer between sources and provides loading for the unused
source in either position.
A rugged formica carrying case is included with each HCS-1
Shunt Set for convenient, dust-free module storage and
transportation. The Model HCS-1 provides primary alternating
current measurement capability with NIST traceability.
See below for the technical specifications of the HCS-1.
|
|
MODEL HCS-1 Primary Shunt Set
TECHNICAL SPECIFICATIONS |
APPLICATIONS:
Any alternating current level is precisely determined by equating it
to the direct current required to produce the same voltage drop
across the HCS-1 Shunt. The DC is accurately measured by use of
a DC Shunt, potentiometer and standard cells. Each of the two
sources must be extremely stable and capable of the required
voltage burdens at rated current. Transfers are accomplished on a
direct 1 for 1 current basis. The measurement accuracy is not a
function of the absolute resistance of the Shunt. The absolute AC
accuracy is a function of the AC-DC difference of the Shunt, the
accuracy of the AC/DC voltage comparison, and the accuracy with
which the DC current can be referenced to a standard cell voltage.
 | Click icon on left to see an illustration showing a typical current measurement circuit employing the HCS-1. |
SPECIFICATIONS:
| Current Ranges | 10,25, 50, 100, and 250 milliamps, each switch selected in one module. Additional individual modules at 500 milliamps, and 1, 2, 5, 10, and 20 amperes. Each module may be used for 50% to 130% of rated current. | Frequency Characteristics | The maximum AC/DC difference at any frequency from 5 Hz to 50 Khz will not exceed 0.02%.
Four terminal input admittance 1MHO/Ampere +/- 1% Option a: 1 MHO/Ampere +/- 0.1%. |
| Dimensions | 17 5/8" W x 12 3/8" H x 4 3/4" D | Accessories | Current transfer switches. Ask for technical bulletin. |
| Weight | Carrying case with 7 modules: 16.5 lbs. Approximate shipping weight: 35 lbs. | | |
Specifications subject to change without notice.
Click to return to top
|
|